Determining the accuracy of manufactured thromboplastin compared to standard thromboplastin in Khartoum state laboratories - Sudan, during April to September 2013
16th World Hematology Congress
March 18-19, 2019 | Rome, Italy

Mohammed Ahmed Mamoun Kissinger

University of Medical Sciences and Technology - Sudan

Posters & Accepted Abstracts: Blood Disord Transfus

Abstract:

Background: PT/INR is used to monitor AVK therapy, Warfarin. There are different thromboplastin reagents and methods used to assess prothrombin status by extrinsic pathway.

Objective: The main aim of this study is to determine the accuracy of manufactured thromboplastin in comparison to standard thromboplastin. The study is laboratory based cross sectional study in which thirty-six laboratories in Khartoum state were evaluated for the performance of PT and INR technique by using certified lyophilized plasma level 1. Data were collected through direct interview done by researcher in which questionnaire was filled and sample were tested in duplicate two different types of thromboplastin one which is used by the laboratories (manufactured thromboplastin) and the other by standard thromboplastin. Data was analyzed by using statistical equation and SPSS. Study shows there were different reagents to measure PT/INR fortress occupy (50%) of them and 8.3% for each one from spectrum, uniplastin, pioplastin, tetlclot, neoplastin, vitro, biomed, spinereact, tico and technoplastin with different methods as manual 2(5.56%), semi automation 28(77.78%) and full automation 1(2.78%). Coagulometers were 15(41.67%) bio-bas (Spinreact) magnetic and O.D 6(16.67%). 21(58.3%) of laboratories were calibrate pipette. Controls used were 20(55.56%) commercial and 14(38.89%) sample. Study were found the CV=0.35, SD±1.12 and mean=1.43 of samples and CV=0.78, SD±1.13 and Mean=3.2 of standards. 14(39%) were correct it with new derived ISI when calculate INR.

Conclusion: There are various reagents used by different laboratories; majority of labs using semiautomated methods and only 2% using automation. There is an inconsistency in the accuracy of the result and different mean, CV and SD between commercial reagent and the control one.

Biography :