Short Communication - (2017) Volume 8, Issue 4
A signal processing approach to reducing noise in the observed transmission through nanohole array sensors used in a microscale calorimeter is presented. The results demonstrate that using rectification and moving average processes, a well-defined Extraordinary Optical Transmission (EOT)-temperature calibration curve can be developed in the presence of a moving light fringe pattern.
Keywords: Signal processing; Moving light fringes; Moving average; Nanohole Array Sensors (NAS); Calorimeter
The use of Nanohole Array Sensors (NHA) to measure the temperature and concentration a change in a microscale calorimeter has been described (Figure 1) [1-5].
Light wave fringes are observed during the reaction process (Figure 2A) involving fluids with different index of refractions [2,3]. These fringes change the incident intensity on the NHA and the observed Extraordinary Optical Transmission (EOT). The induced EOT changes increase the signal noise (Figure 2B) and complicate obtaining an EOT-temperature calibration. Signal processing techniques are used to reduce the EOT noise introduced by these fringe patterns. The EOT noise observed in Figure 2 for the 6 sensors is series of peaks and valleys shortly after the 60 s injection of the ethanol into the water at 400 s. After the injection, the response becomes monotonic because the concentration is uniform throughout the cell and the fringe patterns disappear. Each sensor has a different EOT reading because of slight differences among them (Figure 2).
The first step in the signal processing of the EOT signal is rectifying the data. Each sensor EOT signal of pure water (time <400 s) which is relatively uniform is averaged using one hundred data points starting at 380 s. The rectifying filter compares all the EOT data points of a sensor to its averaged EOT. The points with large EOT value relative to the average are saved. The EOT data with lower value relative to the average will be reflected above the average value with the same intensity. Figure 3 summarizes the raw data and the rectified data for a typical sensor, #24. A moving average filter (35 data points) is also applied to the rectified data to further decrease the EOT noise and better prepare the signal for temperature and concentration calibration.
The temperature of the chamber is measured by a micro thermistor (U.S. SENSOR Corp., USP17640) that is at a similar level in the chamber (Figure 1) as the NHA and experiences similar temperature changes. The result of the final filtered EOT for the 6 sensors in Figure 2 are summarized in Figure 4. The temperature-time trend for the EOT intensity is analogous to that for the thermistor. The EOT variation as a function of the temperature variation is shown in Figure 5 for the injection period for sensor 24. Sensor 24 is in close proximity to the thermistor. Figure 5 is used to calibrate the EOT relative to the temperature.
Figure 5: Temperature-EOT calibration for sensor 24. Dashed linecircle curve is for increasing temperature during dilution as the ethanol mole fraction changes from 0 to 0.103. The region corresponding to cooling at constant concentration results in a monotonic calibration curve (line with square symbols)
The dashed line represents the convolved concentration and temperature signal during the ethanol injection. The solid line is at a constant concentration and is only a function of temperature [1]. After the injection, time greater than 460 s, the chamber cell is cooling at a constant concentration. As observed in Figure 5, the EOT variation as a function of the temperature results in a monotonic relationship that is used for calibration. The results shown in Figure 5 demonstrates the advantages of using the signal processing-rectification method with a moving average to extract more useful information from the EOT measurements than provided by the data in Figure 2.
The consistency between the EOT and temperature measurements is better than Figure 4 that observed in Figure 2.