Research Article - (2015) Volume 6, Issue 6
A sensitive, rapid, accurate and reproducible Isocratic, RP-LC method was developed for determination of Alprazolam and Mebeverine in pharmaceutical dosage form. A Sunfire C18, 5 μm column with mobile phase containing methanol: buffer (0.02M KH2PO4) (70:30). The flow rate was 1.0 ml/min and effluents were monitored at 225 nm with help of PDA detector. The retention time was 6.04 min. for ALP and 3.53 min. for MEB. The linearity was in the range of 0.05-40 μg/ ml for ALP and 0.2-40 μg/ ml for MEB. The proposed method was validated with respect to linearity, accuracy, precision, specificity and robustness. Due to simplicity, rapidity and accuracy of method, believe that the method will be useful for routine quality control analysis.
Alprazolam (ALP) is a white powder and amorphous in nature. Chemically it is 8-chloro-1-methyl-6-phenyl-4H-[1,2,4] triazolo [4,3,- α]-[1,4] benzodiazepine[1] (Figure 1A). It is bind nonspecifically to benzodiazepine receptors BNZ-1, which mediates sleep, and BNZ-2, which affects muscle relaxation, anticonvulsant activity, motor coordination, and memory [2]. Mebeverine (MEB) is white powder. Chemically, it is 3, 4-Dimethoxybenzoic acid 4- [ethyl[2-(4- methoxyphenyl)-1-methylethyl] amino]-butylester and both the drugs are very soluble in water and alcohol [3] (Figure 1B). Mebeverine act on the gut muscles at cellular level to relax them. This relieves painful muscles spasms of gut without affecting its normal motility. It is also an inhibitor of calcium depot replenishment [4]. Therefore it has dual mode of action which normalizes the small motility. The combination of Alprazolam with Mebeverine was useful and safe in irritable bowel syndrome.
Literature survey revealed that alprazolam can be estimated by spectrophotometry [5] and by liquid chromatographic methods [6-11] individually or in combination with other drugs, and mebeverine can be estimated by liquid chromatographic methods individually or in combination with other drugs [12-16]. The reported methods are highly sophisticated, costly, and time-consuming and require special sample preparation. In comparison to LC and other methods, HPLC method is considered to be a good alternative, and it should be widely explored as an important tool in routine drug analysis. UV Spectroscopy method has been reported for simultaneous estimation of both alprazolam and mebeverine in combined dosage form [17].
¨To our knowledge, no article related to the HPLC determination of ALP and MEB in pharmaceutical dosage forms has appeared in the literature. Present study involves development of a high performance thin layer liquid chromatographic method for the determination of ALP and MEB in combination dosage form. A major advantage of HPLC is its ability to analyze several samples simultaneously using a small quantity of mobile phase. This reduces the time and cost of analysis, minimizes exposure risks, and significantly reduces disposal problems of toxic organic solvents, thereby reducing the possibilities of environment pollution.
The aim of the present work was to develop an accurate, repeatable, and specific HPLC method for the determination of ALP and MEB both as a bulk drug and in formulation. The proposed method was validated according to ICH guidelines [18] and its updated international convention.
The liquid chromatographic system consist of Waters series 2998 (Shelton, USA) equipped with a PDA detector, series 515 quaternary isocratic pump and manual injector rheodyne valve with 20 μL fixed loop. The analytes were monitored at 225 nm. Chromatographic analysis was performed on Sunfire C18 column having 250 mm× 4.6 mm i.d. and 5 μm particle size. All the drugs and chemicals were weighed on Mettler Toledo electronic balance (ME204/A04, Mettler Toledo Group).
Analytically pure ALP and MEB were obtained as gift samples from Sun Pharmaceutical Pvt. Ltd., Baroda, India. Methanol (HPLC Grade) SRL Private Ltd. Mumbai, India and Potassium Dihydrogen Ortho Phosphate Buffer – S. D. Fine Pvt. Ltd., Mumbai, India was obtained
Chromatographic conditions
The Sunfire C18 column equilibrated with mobile phase methanol: buffer (0.02M Potassium Dihydrogen Ortho Phosphate Buffer) (70:30) was used. The flow rate was maintained at 1 mL/min, eluent were monitored with PDA detector at 225 nm, and the injection volume was 20 μL. Total run time was kept 8 min.
Preparation of standard stock solutions
Alprazolam and Mebeverine were weighed (10 mg) and transferred to 10 ml volumetric flask containing few ml of methanol. Volume was made up to the mark with methanol to yield a solution containing 1000 μg/ml of alprazolam and mebeverine. Aliquot from the stock solution were appropriately diluted with mobile phase to obtain working standard of 100 μg/ml ALP and MEB.
Method Validation
The method was validated for accuracy, precision, linearity, detection limit, quantitation limit and robustness.
Linearity
Appropriate aliquots of ALP and MEB working standard solution was taken in different 10 mL volumetric flasks and diluted up to the mark with mobile phase to obtain final concentrations of 0.05, 0.5, 5, 10, 20, 40 μg/mL of ALP and 0.2, 0.5, 5, 10, 20, 40 μg/mL of MEB. The solutions were injected using a 20 μL fixed loop system and chromatograms were recorded. Calibration curves were constructed by plotting average peak area versus concentrations and regression equations were computed (n=5).
Precision
The repeatability study was carried out by estimating response six times and results are reported in terms of relative standard deviation. The intra-day and inter-day precision study (intermediate precision) was carried out by estimating the responses 3 times on the same day and on 3 different days for three different concentrations of 0.05, 5, 20 of ALP and 0.2, 5, 20 of MEB and the results are reported in terms of relative standard deviation.
Accuracy
The accuracy of the method was determined by calculating recovery of ALP and MEB by method of standard addition. Known amount of ALP (0, 0.05, 0.1, 0.15 μg/mL), MEB (0, 5, 10, 15 μg/mL) were added to a pre quantified sample solution, and the amount of ALP and MEB were estimated by measuring the peak areas and by fitting these values to the straight-line equation of calibration curve.
Detection limit and quantitation limit
The limit of detection (LOD) is defined as the lowest concentration of an analyte that can reliably be differentiated from background levels. Limit of quantification (LOQ) of an individual analytical procedure is the lowest amount of analyte that can be quantitatively determined with suitable precision and accuracy. LOD and LOQ were calculated using following equation as per ICH guidelines
LOD = 3.3 ×σ /S; LOQ = 10 ×σ /S;
Where σ is the standard deviation of y-intercepts of regression lines and S is the slope of the calibration curve.
Specificity
The specificity was estimated by spiking commonly used excipients (starch, talc and magnesium stearate) into a pre weighed quantity of drug. The chromatogram is taken by appropriate dilutions. Developed spot area and retention time of ALP and MEB were determined and effect of interfering compound was investigated.
Robustness
Robustness of the method was studied by deliberately changing the experimental conditions like flow rate and percentage of organic phase.
Solution stability
Stability of sample solutions were studied at 25 ± 2°C for 24 h.
System suitability
A system suitability test was an integral part of the method development to verify that the system is adequate for the analysis of ALP and MEB to be performed. System suitability test of the chromatography system was performed before validation run. Six replicate injections of a system suitability standard and one injection of a check standard were made. Area, retention time (Rt), capacity factor, column efficiency, symmetric factor, and signal to noise ratio for the five suitability injections were determined.
Analysis of marketed formulation
Twenty tablets were weighed accurately and finely powdered. Tablet powder equivalent to 0.25 mg ALP and 135 mg of MEB were taken in 10 ml volumetric flask. Methanol (5 ml) was added to the above flask and the flask was sonicated for 10 minutes. The solution was filtered using Whatman filter paper No.41 and volume was made up to the mark with the mobile phase.
Appropriate volume of the aliquot was transferred to a 10 ml volumetric flask and the volume was made up to the mark with the mobile phase to obtain a solution containing 0.1 μg/ml of ALP and 54 μg/ml of MEB. The solution was sonicated for 10 min. It was injected as per the above chromatographic conditions and peak area were recorded. The quantification was carried out by keeping this value to the straight line equation of calibration curve.
Optimization of mobile phase
The objective of the method development was to achieve chromatographic peaks for active drug ingredients with less asymmetric factor. The mobile phase methanol: buffer(0.02M Potassium Dihydrogen Ortho Phosphate Buffer) (70:30 %v/v) was found to be satisfactory which gave symmetric peak. Overlaid UV spectra of the drugs showed that ALP and MEB absorbed appreciably at 225 nm (Figure 2), so detection was carried out at 225 nm. The retention time for ALP was 3.53 min. and MEB was 6.04 min (Figure 3). The asymmetric factor for ALP and MEB was 0.45 and 0.57, respectively. The mobile phase flow rate was maintained at 1 mL/min. Overlain chromatograms of ALP and MEB were shown in Figure 4.
The calibration curve for ALP and MEB was found to be linear in the range of 0.05-40 μg/ml and 0.2-40 μg/ml with a correlation coefficient of 0.995 and 0.999. The summary of validation and system suitability parameters were shown in Table 1.
Parameters | ALP | MEB |
---|---|---|
Linearity ( µg/ml) | 0.05-40 | 0.2-40 |
Retention Time (min) | 6 | 3.5 |
Detection limit ( µg/ml) | 0.08 | 0.01 |
Quantitation limit ( µg/ml) | 0.24 | 0.05 |
Accuracy (%) | 99.42-100.55 | 98.29-101.39 |
Precision (%RSD) | ||
Intra-day (n = 3) | 0.70-1.31 | 0.69-1 |
Inter-day (n = 3) | 0.82-1.81 | 1.14-1.82 |
Specificity | Specific | Specific |
Robustness | Robust | Robust |
Solvent suitability | Suitable for 24 hrs. | Suitable for 24 hrs. |
Retention time(min) | 6.04 | 3.53 |
Theoretical Plates | 5053 | 6903 |
Asymmetric factor | 0.45 | 0.57 |
Resolution | 2.57 |
Table 1: Summary Of Validation And System Suitability Parameters
The intra-day and inter-day precision studies were carried out. The low RSD value indicate that the method is precise. The accuracy of the method was determined by calculating recovery of ALP and MEB by method of standard addition. The percentage recovery was found to be 98.29–101.39 for ALP and 99.42 - 100.55 for MEB, respectively. This values indicate that the method is accurate (Table 2).
Level |
Amount Added (µg/ml) | Amount Recovered (µg/ml) (n=3) | % Recovered ± S.D | |||
---|---|---|---|---|---|---|
MEB | ALP | MEB | ALP | % MEB | % ALP | |
0 | 10 + 0 | 0.1+0 | 10.05 | 0.10 | 100.55 ± 1.08 | 101.39 ± 4.06 |
50 | 10+5 | 0.1+0.05 | 14.99 | 0.14 | 99.98 ± 1.44 | 98.29 ± 3.92 |
100 | 10+10 | 0.1+0.1 | 19.94 | 0.20 | 99.42 ± 1.51 | 100.49 ± 5.97 |
150 | 10+15 | 0.1+0.15 | 24.99 | 0.24 | 99.96 ± 1.51 | 99.23 ± 7.58 |
Table 2: Accuracy Study Of Alp And Meb By Proposed Rp-Lc Method.
The detection limit for ALP and MEB was found to be 0.080 μg/mL and 0.016 μg/mL, respectively, while quantitation limit was found to be 0.243 μg/mL and 0.050 μg/mL, respectively The above data shows that a nano gram quantity of the drug can be accurately and precisely determined. Robustness study was performed by deliberately changing the experimental conditions like flow rate from 1 ml/min to 0.8 ml/min and 1.2 ml/min. The composition of mobile phase was changed varying the proportion of methanol by 2 %. In both the conditions the recovery of the drug was determined and the RSD was found to be less than 2% (Table 3). Stability of standard and sample solution of ALP and MEB were evaluated at room temperature for 24 hr. The drug was found to be stable with a amount of drug found more than 98%.
Method Parameter | Normal Condition | Deliberate Change | %RSD of peak area (n=3) | |
---|---|---|---|---|
ALP | MEB | |||
Flow rate | 1.0 ml/min | 0.8 ml/min | 0.49 | 1.14 |
1.2 ml/min | 0.60 | 1.06 | ||
Mobile phase ratio | Methanol: Buffer (70:30) | 72:28 | 1.01 | 1.90 |
68:32 | 1.12 | 1.81 | ||
pH of mobile phase ratio | 5.0 | 5.2 | 0.58 | 0.67 |
4.8 | 0.69 | 1.02 |
Table 3: Robustness Data For Alp and Meb By Proposed Rp- Lc Method.
The proposed method was successfully applied to the determination of ALP and MEB in its dosage form. The % recovery was found to be 100.51 10.58 and 100.55 ± 1.08, respectively which was comparable with the labeled amount (n=3) (Table 4). The chromatogram of marketed formulation was shown in Figure 5.
Formulation MEBASPA- AL |
Amount drug taken (mg/mL) | Amount of drug found | % Drug found Mean ± SD* (n =3) |
|||
---|---|---|---|---|---|---|
ALP | MEB | ALP | MEB | ALP | MEB | |
0.1 | 10 | 0.105 | 10.05 | 100.51±10.58 | 100.55±1.08 |
Table 4: Analysis of Marketed Formulation.
Proposed study describes RP-LC method for the simultaneous quantification of ALP and MEB in its formulation. The method was validated and found to be simple, sensitive, accurate and precise. Statistical analysis proved that method was repeatable and selective for the simultaneous analysis of ALP and MEB without any interference from the excipients. The method was successfully used for determination of drug in their formulation.
The authors are very thankful to Principal, Indukaka Ipcowala College of Pharmacy, New Vallabh Vidyanagar for providing necessary facilities to carry out research work.